Mass spectrometry
peak picker
Measuring a chromatographic peak involves choosing the intended signal and integrating its area above the baseline. I built this LC–MS tool to test those steps separately and give analysts a way to review the result.
- My role
- Developer
- Medium
- LC–MS data analysis
The simulated trace peaks at 97 seconds. Moving either integration boundary inward excludes some of the signal above the baseline.
Area error and peak selection
Across 348 comparable simulated single peaks, the revised integration method reduced median absolute area error from 4.19% to 1.21%.
In the separate selection test, the tool chose the intended peak in 401 of 576 simulated windows. It chose a different peak in 136 windows and left 39 unresolved. Those results are why the review step shows the selected signal, estimated baseline, and integration boundaries before the analyst exports a result.