Mass spectrometry
peak picker

Measuring a chromatographic peak involves choosing the intended signal and integrating its area above the baseline. I built this LC–MS tool to test those steps separately and give analysts a way to review the result.

By
My role
Developer
Medium
LC–MS data analysis

The simulated trace peaks at 97 seconds. Moving either integration boundary inward excludes some of the signal above the baseline.

Testing the integration window

The shaded region shows the area above the estimated baseline within the two boundaries. Move either boundary toward 97 seconds to see how the integrated area changes.

Compare the original plot
Version 1.2 chromatogram of the same synthetic peak at 97 seconds.
The same synthetic trace in the version 1.2 release.

Area error and peak selection

Across 348 comparable simulated single peaks, the revised integration method reduced median absolute area error from 4.19% to 1.21%.

In the separate selection test, the tool chose the intended peak in 401 of 576 simulated windows. It chose a different peak in 136 windows and left 39 unresolved. Those results are why the review step shows the selected signal, estimated baseline, and integration boundaries before the analyst exports a result.

For two peaks in one public marine sample, alternative areas differed from reference by −0.185% and −0.262%. The original differences were +0.762% and −3.326%.
Two peaks from one public marine injection, compared with published Skyline reference areas. Source study